Parameter | Value |
|---|---|
Measures | |
Measure principle | Conductivity sensor with 4 electrodes (2 graphic, 2 platinum) |
Measure ranges conductivity | 0-200,0 µS/cm • 0-2000 µS/cm • 0,00-20,00 mS/cm • 0,0-200,0 mS/cm |
Resolution | 0,01 to 1 according the range |
Accuracy | +/- 1% of the full range • Beyond 100 mS/cm use appropriate buffer solution |
Measure range salinity | 5-60 g/Kg |
Measure range TDS-KCl | 0-133 000 ppm |
Response time | < 5 s |
Working temperature | 0°C to +50°C |
Temperature compensation | NTC |
Stocking temperature | -10°C to +60°C |
Signal interface | Modbus RS-485 (standard) and SDI-12 (option) |
Maximum refreshing time | Max < 1 s |
Sensor power-supply | 5 to 12 volts |
Electric consumption | Standby: 25 µA • Average RS485 (1 measure/seconde): 6,3 mA / Average SDI12 (1 measure/seconde): 9,2 mA • Current pulse: 500 mA |
Sensor | |
Dimensions | Diameter: 27 mm; Length: 157 mm |
Weight | 300 g (sensor + cable 3 meters) |
Material | PVC, DELRIN, stainless steel |
Maximum pressure | 5 bars |
Connection | 9 armoured connectors, polyurethane jacket, bare-wires or waterproof Fisher connector |
Protection | IP68 |