Measures | |
|---|---|
Measure principle | Optical measure by luminescence |
Measure ranges | 0,00 to 20,00 mg/L / 0,00 to 20,00 ppm / 0-200% |
Resolution | 0,01 |
Accuracy | +/- 0,1 mg/L, +/- 0,1 ppm, +/- 1% (+/-5% if EMI Perturbations are more than 10V/m) Range 0-100% — The sensor must be fully immerged to ensure maximum accuracy |
Limit of detection | 0.7% |
Limit of quantification | 2.2% |
Repeatability (100% Sat) | 0.2% |
Linearity | >0.99 |
Response time | 0 -> 100%; T90 < 40s / 100 -> 0%; T90 < 65s |
Frequency of recommended measure | >5s |
Water Flow | No movement required for measurement |
Temperature compensation | Via NTC |
Temperature | 0.00-50.00 °C / Accuracy: +/- 0.5 °C |
Stocking temperature | -10°C to +60°C |
Temperature range | 0°C to 50°C |
Temperature accuracy | +/- 0,5°C |
Communication Power Supply | |
|---|---|
Signal interface | Modbus RS-485 or SDI-12 — 1,2: The sensor responds in Modbus / SDI12 including during Standby; 3: The use and connexion of SDI12 bus may increase the standby power consumption* up to 100uA depending the level of the line (high or low). The consumption is not increased if the SDI12 line is disconnected or released to 0V (Modbus RTU only) |
Sensor power-supply (RS485 and SDI12) | 5V to 12V DC (warm-up time 100 ms) — 1: Absolute minimum 4.5V with 1m of cable, boot and precision not guaranteed under 5V; 2: Minimum voltage subjected to cable length-related losses; 3: 13V Absolute maximum with a more than 2mA continuous over consumption; 4: Small overconsumption between 12V and 12.5V |
Consumption | Standby: 25 µA / Average RS485 (1 meas/s): 3.2 mA / Average SDI12 (1 meas/s): 6 mA / Current pulse: 85 mA (3 mS) — For more details, refer to the user manual |
Sensor | |
Dimensions | Standard version: Diameter: 27mm; Length: 143mm / More protective strainer: Diameter: 27mm; Length: 166mm |
Weight | 300 g (sensor + cable 3 m) |
Material | Black POM C, PVC |
Maximum pressure | 5 bars |
Connection | 9 armoured connectors, polyurethane jacket, bare-wires |
Ingress Protection rating | IP68 |