Replacement Interval

Sensor Blade Replacement
It is recommended to replace the sensor blade's rubber insert every quarter. Follow the steps below:

  1. Locate the blade on the sensor, as shown in the diagram on the left.

  2. Using the appropriate screwdriver, remove the screws securing the blade.

  3. Remove the blade and take out the rubber insert from inside. Replace it with a new rubber insert.

  4. Reattach the blade, then tighten and secure the screws firmly with the screwdriver.


Calibration Interval

Cleaning Interval

Supmea

SUP-PSS200

The Suspended Solids Meter uses infrared scattered light absorption technology in combination with the ISO 7027 method to provide continuous and accurate measurement of suspended solids and sludge concentration.

Based on the ISO 7027 standard, the dual infrared scattered light technology ensures that measurements of suspended solids and sludge concentration are not affected by the color or turbidity of the sample. An optional self-cleaning function is available to suit different operating environments.

Property

Details

Main materials

Main Body: SUS316L (Ordinary Version), Titanium Alloy (Seawater Version)


Upper and Lower Cover: PVC; Cable: PVC

Waterproof rate

IP68/NEMA6P

Storage temperature

-15~65°C

Operation Temperature

-15~60°C

Output

three-way analog output 4-20mA, response parameters and corresponding scope can be programmed Note: the maximum load is 500 ohms

Relay

three-way relay can be set up, and response parameters and response values can be programmed

Digital Communication

MODBUS RS485 communication function, which can transmit real-time measurements

Power Supply

AC: AC220V, 50HZ, 5W  DC: DC24V

Size

145125162mm LWH

Weight

1.3KG

Warranty

1 Year

Cable Length

Standard 10-Meter Cable, Max Length: 100 Meters