Replacement Interval

Sensor Blade Replacement
It is recommended to replace the sensor blade's rubber insert every quarter. Follow the steps below:
Locate the blade on the sensor, as shown in the diagram on the left.
Using the appropriate screwdriver, remove the screws securing the blade.
Remove the blade and take out the rubber insert from inside. Replace it with a new rubber insert.
Reattach the blade, then tighten and secure the screws firmly with the screwdriver.
Calibration Interval
Cleaning Interval
The Suspended Solids Meter uses infrared scattered light absorption technology in combination with the ISO 7027 method to provide continuous and accurate measurement of suspended solids and sludge concentration.
Based on the ISO 7027 standard, the dual infrared scattered light technology ensures that measurements of suspended solids and sludge concentration are not affected by the color or turbidity of the sample. An optional self-cleaning function is available to suit different operating environments.
Property | Details |
|---|---|
Main materials | Main Body: SUS316L (Ordinary Version), Titanium Alloy (Seawater Version) |
Upper and Lower Cover: PVC; Cable: PVC | |
Waterproof rate | IP68/NEMA6P |
Storage temperature | -15~65°C |
Operation Temperature | -15~60°C |
Output | three-way analog output 4-20mA, response parameters and corresponding scope can be programmed Note: the maximum load is 500 ohms |
Relay | three-way relay can be set up, and response parameters and response values can be programmed |
Digital Communication | MODBUS RS485 communication function, which can transmit real-time measurements |
Power Supply | AC: AC220V, 50HZ, 5W DC: DC24V |
Size | 145125162mm LWH |
Weight | 1.3KG |
Warranty | 1 Year |
Cable Length | Standard 10-Meter Cable, Max Length: 100 Meters |